The Most Advanced Plasma Sterilizer

LOWTEM Crystal 120D

  • 1. Rapid Warmup & Dry System
    2. Convenient and Safe Sterilizing Agent
    3. 7” full touch LCD, Easy Monitoring the Cycle Information
    4. Plug and Play
    5. USB History Memory
    6. Automatic PM Alarm System
    7. Built in Thermal Printer, Printout with Actual Cycle Information
    8. Login Function
    9. Monitoring System
    10. Auto Interlock & Open System
    11. Double Door System

This product is a medical device, please read the precautions and instructions for use carefully.

Product Features

  • 1Double Door System

    LOWTEM Crystal 120D has an optimizing pass-through double door system for CSSD, satisfying a need for separation of clean and sterile zone.

  • 2Auto Interlock & Open System

    One gentle push makes the chamber door locked, and the chamber door can be opened automatically when the door sensor detects finger movement. LOWTEM Series’ extraordinary user friendly functions provide convenient and safe environment.

  • 3Monitoring System

    1.Cycle temperature, pressure, and sterilization cycle status can be remotely monitored.
    2.Thanks to the monitoring system, LOWTEM Series can be maintained and upgraded easily.

  • 4Login Function

    LOWTEM Series provide User ID and PW, so that hospital can manage users who access the sterilizer.

  • 5Built in Thermal Printer, Printout with Actual Cycle Information

    Built-in thermal printer uses 60mm (Ø) Roll paper. Users do not need to replace it often.
    The printout from this printer provides temperature, pressure with the cycle graph on the paper.

  • 6Automatic PM Alarm System

    LOWTEM Series provide alarms to prevent from irregular maintenance.
    Automatic PM Alarm System is very useful for users and engineers to be noticed.

  • 7USB History Memory

    LOWTEM Series’ sterilization history can be downloaded to your USB memory stick, and maintained at your convenience. Furthermore, it can be remote-controlled by its own out-of-state network system.

  • 8Plug and Play

    LOWTEM Series’ warming up time is very rapid (Maximum 15 min) after main power is on, so that standby power consumption is zero during off-duty or overnight.

  • 97” full touch LCD, Easy Monitoring the Cycle Information

    1.LOWTEM Series’ 7”full touch LCD monitor provides with all functions of the machine using simple letters, pictures, and icons.
    2.User can monitor the actual temperature and pressure from the graph on the monitor.

  • 10Convenient and Safe Sterilizing Agent

    1.“Tank type” sterilant is safe for users to store, to deliver, and to use.
    2.The automatic system for changing and installing sterilant tank is convenient and safe for users.
    3.LOWTEM Series’ tank type sterilant does not need to be replaced often, as it can be used scores of times.

  • 11Rapid Warmup & Dry System

    1.The world’s first patented “Rapid Warmup and Dry System” allows to overcome the humidity-related problem.
    2.LOWTEM Series automatically removes residual moisture through the enhanced drying performance for medical devices in warm up phase.
    3.“Rapid Warmup and Dry System” gets rid of Cold Points and maintains same temperature inside chamber, which is very effective for sterilizing the complicated medical devices with its strong penetration power.

Product specification

Item Description

Sterilization

Hydrogen Peroxide
20 Cycles / Bottle
Total Cycle Time Quick

30±5 Min

Standard

40±5 Min

Special

50±5 Min

Cycle Temperature

55±5℃

SAL (Sterility Assurance Level) 10-6
By-Products Oxygen And Vapor Water Only
Lumen Claims Rigid Lumen & Flexible Lumen, Endoscopes
Chamber Type

Double Door SystemRectangular

Material Stainless Steel(SUS)
Dimensions Overall

680mm(W) x 1,575mm(H) x 784mm(D)

Chamber

420mm(W) x 400mm(H) x 720mm(D)

Volume Total : 130 Liter
Usable : 117 Liter (420mm(W) x 400mm(H) x 720mm(D))
Weight 300kg
Control Microprocessor & Windows CE Embedded
Cycle Information

USB

PM Cycle Automatic Alarm & Setup
Electrical

220-230V, 50/60Hz, 1 Phase, 3000W

Installation Requirements

Front : 100cm
Rear  : 100cm

Left Side, Right Side : 10cm
Placement : Built-in Wheels Provide Mobility

Room Conditions 5 ~ 40℃, 0 ~ 95%RH(Non-Condensing)
Printer Built-in Thermal Printer(60Ø Roll paper)
Cycle Parameters (Temp, Pressure, Time, Daily&Total Cycle, Etc)
Alarm & Error Display
Others Emergency Stop(Front), Operator ID Login, Self-Diagnosis,
Multi-Language, RFID Coding System